EndoSyncPlus Overview Part 3 of 3

Allen Ali Nasseh   uploaded 1 year ago

In this video, Dr. Nasseh shares some of the specific memory settings as it relates to safety and efficiency and makes some clinical recommendations for specific settings

Comments

  • This topic has 2 replies, 2 voices, and was last updated 1 year, 1 month ago by Allen Ali NassehAllen Ali Nasseh.
Viewing 1 reply thread
  • Author
    Posts
    • #3436
      Avatarjeffthedentist

      Is there a way to run Wave one gold to auto reciprocate? Or just setting torque to 0.2?

    • #3497
      Allen Ali NassehAllen Ali Nasseh

      @Jeffthedentist: yes, you can do the same thing with WOG. But you need to use EndoSync Plus, not the original EndoSync as you will need Reverse OTR with 0.2Ncm setting with a cutting angle of 150 and non-cutting angle of 30.
      Hope this helps.
      Cheers!